This will return an integer followed by a space, carriage return and a line feed character
(ASCII 10).
Example - the following command would retrieve the current Cycles setting:
$CYCLES ?
The pulse generator will return the following:
Echo Enabled: CYCLES ?<cr><lf>
100<cr><lf>
Echo Disabled: 100<cr><lf>
Communication Protocol
Communication parameters are set at 9600 baud, 8 bits, parity = none, and 1 stop bit. To help
establish RS232 communications, the pulse generator emits the characters QC approximately
every 200 ms after power up and until the 9314E receives a $. Both the GPIB and the
RS232 interface may be used at the same time. The instrument will respond to the port which
sent the command.