19
9314E Communication Commands
Command Name Parameters Description
MODE 0 = Continuous mode
1 = Burst mode
2 = Duty Cycle mode
3 = Single Shot mode
4 = External Trigger mode
Sets the pulse generator mode.
GATE 0 = off
1 = active low
2 = active high
Enables the external trigger input to function as a gate
controlling all outputs. Cannot be used with external
trigger mode.
ToPer 1000 - 999999998 Sets the To internal sync period, in microseconds * 10.
DC:On 1 - 10000 Sec Sets the on time, in seconds, for the Duty Cycle mode.
DC:Off 1 - 10000 Sec Sets the off time, in seconds, for the Duty Cycle mode.
TRIG 0 = falling edge
1 = rising edge
Sets the active edge of the Ext/Gate signal when used as
an external trigger.
CYCLES 0 - 10000 cycles Sets the number of cycles for the Duty Cycle mode.
#/BURST 1 - 30000 pulses Sets the number pulses in the Burst mode.
T1a -999999998 - +999999998 Sets the delay, in microseconds * 10, for edge 1a.
T1a= 0 - 8
( 0=To, 1=1a, 2=1b, etc.)
Sets the sync edge for edge 1a.
T1b -999999998 - +999999998 Sets the delay, in microseconds * 10, for edge 1b.
T1b= 0 - 8
( 0=To, 1=1a, 2=1b, etc.)
Sets the sync edge for edge 1b.
T1:Pol 0 = Off
1 = Negative (active low)
2 = Positive (active high)
Sets the polarity of the output for Channel 1.
T2a -999999998 - +999999998 Sets the delay, in microseconds * 10, for edge 2a.
T2a= 0 - 8
( 0=To, 1=1a, 2=1b, etc.)
Sets the sync edge for edge 2a.
T2b -999999998 - +999999998 Sets the delay, in microseconds * 10, for edge 2b.
T2b= 0 - 8
( 0=To, 1=1a, 2=1b, etc.)
Sets the sync edge for edge 2b.
T2:Pol 0 = Off
1 = Negative (active low)
2 = Positive (active high)
Sets the polarity of the output for Channel 2.
T3a -999999998 - +999999998 Sets the delay, in microseconds * 10, for edge 3a.
T3a= 0 - 8
( 0=To, 1=1a, 2=1b, etc.)
Sets the sync edge for edge 3a.
T3b -999999998 - +999999998 Sets the delay, in microseconds * 10, for edge 3b.
T3b= 0 - 8
( 0=To, 1=1a, 2=1b, etc.)
Sets the sync edge for edge 3b.